Alat uji laju transmisi uap air (WVTR) metode gravimetri satu spesimen untuk film plastik, film komposit, lembaran, dan bahan lainnya, sesuai ASTM E96.
| Specification | Value |
|---|---|
| Specification (W301) | |
| Test Range | 0.1 ~ 10,000 g/m²·24h (film test); 0.0003318 ~ 33.18 g/pkg·24h (container test) |
| Resolution | 0.0001 g |
| Number of Specimens | 1 |
| Temperature Range | Ambient ~ 55°C (standard) |
| Temperature Accuracy | ±0.1°C |
| Humidity Range | 10%RH ~ 50%RH |
| Humidity Accuracy | ±1%RH |
| Test Area | 33.18 cm² |
| Sample Thickness | ≤3 mm (customization available) |
| Sample Size | Φ72 mm |
| Instrument Dimension | 440 mm (W) x 560 mm (D) x 510 mm (H) |
| Power Supply | AC 100-240V 50/60Hz (can be changed to local requirement) |
| Net Weight | 42 kg |
| Test Standards | ASTM E96, ASTM D1653, ISO 2528, TAPPI T464, DIN 53122-1, JIS Z0208, GB/T 1037, GB/T 16928, YBB 00092003 |
| Configuration | |
| Standard | Instrument, micro printer, professional software, network cable, test dishes, gas desiccant device, sample cutter, syringe |
| Optional | Computer, reference film, desiccant container, 4A molecular sieve, container testing device |